Introduction
Lyser is a specialized chemical solution used to dye cellular components (e.g., nuclei, cytoplasm) in blood samples, enabling automated hematology instruments to: differentiate blood cell types (WBCs, RBCs, platelets), identify morphological abnormalities, quantify cellular parameters. Lyser for Sysmex hematology instrument is specially for Sysmex series hematology analyzers, including different models of 5-part and 3-part analyzer series.
Specifications
- For Sysmex 5-part series:
XN series B1/B2/B3/B4
|
Code |
Description |
Specification |
Period of Validity |
Temperature |
|
0307004 |
STAC-LYSER-SYSSLS |
1.5Lx2bottles/box |
24m |
2-35℃ |
|
0307005 |
STAC-LYSER-SYSWNR |
4Lx2bottles/box |
24m |
2-35℃ |
|
0307006 |
STAC-LYSER-SYSWDF |
4Lx2bottles/box |
24m |
2-35℃ |
|
0307007 |
STAC-LYSER-SYSWPC |
1.5Lx2bottles/box |
24m |
2-35℃ |
|
030700401 |
STAC-LYSER-SYSSLS |
500mlx3bottles/box |
24m |
2-35℃ |
|
030700501 |
STAC-LYSER-SYSWNR |
5L/box |
24m |
2-35℃ |
|
030700601 |
STAC-LYSER-SYSWDF |
5L/box |
24m |
2-35℃ |
- For Sysmex 5-part series:
XT-1800i/2000i/4000i
|
Code |
Description |
Specification |
Period of Validity |
Temperature |
|
030403 |
STAC-LYSER-SYS-FBA |
5L/box |
24m |
2-35℃ |
|
030003 |
STAC-LYSER-SYS-SLS |
5L/box |
24m |
2-35℃ |
|
030501 |
STAC-LYSER-SYS-4DL |
5L/box |
24m |
2-35℃ |
|
030502 |
STAC-LYSER-SYS-4DS |
42mlx3units/box |
24m |
2-35℃ |
XE-2100/XE-5000
|
Code |
Description |
Specification |
Period of Validity |
Temperature |
|
030403 |
STAC-LYSER-SYS-FBA |
5L/box |
24m |
2-35℃ |
|
030003 |
STAC-LYSER-SYS-SLS |
5L/box |
24m |
2-35℃ |
|
030501 |
STAC-LYSER-SYS-4DL |
5L/box |
24m |
2-35℃ |
|
030502 |
STAC-LYSER-SYS-4DS |
42mlx3units/box |
24m |
2-35℃ |
|
030504 |
STAC-LYSER-SYS-IM |
10L/box |
24m |
2-35℃ |
|
030505 |
STAC-LYSER-SYS-NR&Dye |
1L+12ml/box |
18m |
2-35℃ |
XS-500i/XS800i/XS900i/XS1000i
|
Code |
Description |
Specification |
Period of Validity |
Temperature |
|
030003 |
STAC-LYSER-SYS-SLS |
5L/box |
24m |
2-35℃ |
|
030501 |
STAC-LYSER-SYS-4DL |
5L/box |
24m |
2-35℃ |
|
030502 |
STAC-LYSER-SYS-4DS |
42ml x 3 units/box |
24m |
2-35℃ |
- For Sysmex 3-part series:
KX-21/KX-21N/poch-80i/poch-100i/XP-100/XP-300
|
Code |
Description |
Specification |
Period of Validity |
Temperature |
|
030301 |
STAC-LYSER-SYS-WH |
500mlx 3bottles/box |
24m |
2-35℃ |
K-1000/K-4500
|
Code |
Description |
Specification |
Period of Validity |
Temperature |
|
030003 |
STAC-LYSER-SYS-SLS |
5L/box |
24m |
2-35℃ |
|
030201 |
STAC-LYSER-SYS-3WP |
5L/box |
24m |
2-35℃ |
Product detail
Application: Lyser for Sysmex hematology instrument is applicable to the Sysmex 5-part series: XN-B1/XN-B2/XN-B3/XN-B4,XT-1800i/2000i/4000i,XE-2100/XE-5000,XS-500i/XS800i/XS900i/XS1000i and Sysmex 3-part series:KX-21/KX-21N/poch-80i/poch-100i/XP-100/XP-300, K-1000/K-4500 hematology analyzer line.
- Origin: Shanghai, China
- Certification: CE, ISO 13485, CNAS L18717, MSDS
- Price: Negotiable
- Delivery date: Negotiable
- Packing: export standard
- MOQ: 1 set
- Cost advantage: Compared with the original factory reagents, it has great cost effectiveness.
FAQ
Q: What causes "Lyser Empty" alarm (Code 103000)?
A: Solution:
Replace depleted lyser bottle.
Prime reagent line: Maintenance > Reagent Prime > Select Lyser.
If persists, check bottle sensor alignment (clean with alcohol swab).
Q: Why does HGB show high blank (>0.2 g/dL)?
A: Causes:
● Lyser contamination (e.g., dust in nozzle)
● Expired reagent (verify batch expiration)
● Crystallization in pipes (common in humidity <30%)
Fix:

Q: Shelf life after opening?
A: Storage Temp: Maximum Duration
15-25°C : 90 days
>30°: 30 days
Note: Write opening date on bottle!
Hot Tags: lyser for sysmex hematology instrument, China lyser for sysmex hematology instrument manufacturers, suppliers, Cellsheath for Sysmex Hematology Instrument, Cleaner for Sysmex Hematology Instrument, Diluent for Sysmex Hematology Instrument, Dye for Sysmex Hematology Instrument, Lyser for Sysmex Hematology Instrument, reagents for sysmex series